Ormar za ispitivanje temperature ex zaštićenCTS
T-30/950 Ex/I
Ormar za ispitivanje temperature ex zaštićen
CTS
T-30/950 Ex/I
godina proizvodnje
2011
Stanje
Polovno
Lokacija
Borken 

Slike prikazuju
Prikaži kartu
Podaci o mašini
- Naziv mašine:
- Ormar za ispitivanje temperature ex zaštićen
- Proizvođač:
- CTS
- Model:
- T-30/950 Ex/I
- Godina proizvodnje:
- 2011
- Stanje:
- vrlo dobro (rabljeno)
Cijena i lokacija
- Lokacija:
- Einsteinstraße 8a, 46325 Borken, Deutschland

Pozovite
Detalji ponude
- ID oglasa:
- A7538378
- Referentni broj:
- 23117
- Ažuriranje:
- posljednji put 27.06.2025
Opis
CTS
Type: T-30/950 Ex/I
CTS Temperature Test Chamber, Explosion-Proof
Control – 32-bit controller
LCD control panel
Digital display for set and actual values
Numerical input of set values
Error messages in plain text
RS 232 interface
Free switching channels for product-side test specimen activation
Temperature Testing:
Temperature range: -30°C to +80°C
Hnsdeihb Amspfx Af Asw
Temperature deviation: ±0.3 K over time
Temperature change rate:
Heating: 2 K/min
Cooling: 1.5 K/min down to -10°C
Thermal radiation: approx. 1000 W at +20°C
approx. 200 W at -25°C
Nitrogen inerting with residual oxygen monitoring
With nitrogen inerting, the test chamber is first purged with a large flow rate of N2 until the
o2 content drops below 3%. Then, it switches to a base flow amount (approx. 10% of the large flow rate).
The nitrogen flow for the base amount is measured and displayed using a flowmeter and monitored via a limit switch.
O2 monitoring continuously extracts gas from the test chamber via a flame arrestor and passes it over an O2 sensor which checks the residual oxygen content of the extracted air.
Afterwards, the gas is returned to the test chamber via a second flame arrestor.
The extraction pump is monitored by a volumetric flow meter. Additional details can be found in the Dräger user manual in the appendix.
The gas system is currently deactivated!
All technical data refer to an ambient temperature of +25°C, nominal voltage 400V, without test specimen, without irradiation, and without additional equipment.
Dimensions:
Test chamber volume: 950 liters
Test chamber dimensions: Height approx. 700 mm / Width approx. 900 mm / Depth approx. 1500 mm
External dimensions: Height approx. 2400 mm / Width approx. 1270 mm / Depth approx. 3125 mm
Operating data:
Electrical connection: 400V +6/-10 %, 3/N, 50 Hz
Nominal power: 21.6 kW
Rated current: 36.5 A
Protection class: IP 54
Condensate drain: non-pressurized, R 1/2" inner thread
Sound pressure level:
Oglas je automatski preveden. Moguće su greške u prijevodu.
Type: T-30/950 Ex/I
CTS Temperature Test Chamber, Explosion-Proof
Control – 32-bit controller
LCD control panel
Digital display for set and actual values
Numerical input of set values
Error messages in plain text
RS 232 interface
Free switching channels for product-side test specimen activation
Temperature Testing:
Temperature range: -30°C to +80°C
Hnsdeihb Amspfx Af Asw
Temperature deviation: ±0.3 K over time
Temperature change rate:
Heating: 2 K/min
Cooling: 1.5 K/min down to -10°C
Thermal radiation: approx. 1000 W at +20°C
approx. 200 W at -25°C
Nitrogen inerting with residual oxygen monitoring
With nitrogen inerting, the test chamber is first purged with a large flow rate of N2 until the
o2 content drops below 3%. Then, it switches to a base flow amount (approx. 10% of the large flow rate).
The nitrogen flow for the base amount is measured and displayed using a flowmeter and monitored via a limit switch.
O2 monitoring continuously extracts gas from the test chamber via a flame arrestor and passes it over an O2 sensor which checks the residual oxygen content of the extracted air.
Afterwards, the gas is returned to the test chamber via a second flame arrestor.
The extraction pump is monitored by a volumetric flow meter. Additional details can be found in the Dräger user manual in the appendix.
The gas system is currently deactivated!
All technical data refer to an ambient temperature of +25°C, nominal voltage 400V, without test specimen, without irradiation, and without additional equipment.
Dimensions:
Test chamber volume: 950 liters
Test chamber dimensions: Height approx. 700 mm / Width approx. 900 mm / Depth approx. 1500 mm
External dimensions: Height approx. 2400 mm / Width approx. 1270 mm / Depth approx. 3125 mm
Operating data:
Electrical connection: 400V +6/-10 %, 3/N, 50 Hz
Nominal power: 21.6 kW
Rated current: 36.5 A
Protection class: IP 54
Condensate drain: non-pressurized, R 1/2" inner thread
Sound pressure level:
Oglas je automatski preveden. Moguće su greške u prijevodu.
Dobavljač
Napomena: Registrujte se besplatno ili prijavite, za pristup svim informacijama.
Registriran od: 2012
Pošaljite upit
Telefon & Faks
+49 2861 ... oglasi
Vaš oglas je uspješno izbrisan
Došlo je do greške

















