CTS
Type: T-30/950 Ex/I
CTS Temperature Test Chamber, Explosion-Proof
Control – 32-bit controller
LCD control panel
Digital display for set and actual values
Hedpfxsihb Ams Aafog
Numerical input of set values
Error messages in plain text
RS 232 interface
Free switching channels for product-side test specimen activation
Temperature Testing:
Temperature range: -30°C to +80°C
Temperature deviation: ±0.3 K over time
Temperature change rate:
Heating: 2 K/min
Cooling: 1.5 K/min down to -10°C
Thermal radiation: approx. 1000 W at +20°C
approx. 200 W at -25°C
Nitrogen inerting with residual oxygen monitoring
With nitrogen inerting, the test chamber is first purged with a large flow rate of N2 until the
o2 content drops below 3%. Then, it switches to a base flow amount (approx. 10% of the large flow rate).
The nitrogen flow for the base amount is measured and displayed using a flowmeter and monitored via a limit switch.
O2 monitoring continuously extracts gas from the test chamber via a flame arrestor and passes it over an O2 sensor which checks the residual oxygen content of the extracted air.
Afterwards, the gas is returned to the test chamber via a second flame arrestor.
The extraction pump is monitored by a volumetric flow meter. Additional details can be found in the Dräger user manual in the appendix.
The gas system is currently deactivated!
All technical data refer to an ambient temperature of +25°C, nominal voltage 400V, without test specimen, without irradiation, and without additional equipment.
Dimensions:
Test chamber volume: 950 liters
Test chamber dimensions: Height approx. 700 mm / Width approx. 900 mm / Depth approx. 1500 mm
External dimensions: Height approx. 2400 mm / Width approx. 1270 mm / Depth approx. 3125 mm
Operating data:
Electrical connection: 400V +6/-10 %, 3/N, 50 Hz
Nominal power: 21.6 kW
Rated current: 36.5 A
Protection class: IP 54
Condensate drain: non-pressurized, R 1/2" inner thread
Sound pressure level:
Oglas je automatski preveden. Moguće su greške u prijevodu.
Type: T-30/950 Ex/I
CTS Temperature Test Chamber, Explosion-Proof
Control – 32-bit controller
LCD control panel
Digital display for set and actual values
Hedpfxsihb Ams Aafog
Numerical input of set values
Error messages in plain text
RS 232 interface
Free switching channels for product-side test specimen activation
Temperature Testing:
Temperature range: -30°C to +80°C
Temperature deviation: ±0.3 K over time
Temperature change rate:
Heating: 2 K/min
Cooling: 1.5 K/min down to -10°C
Thermal radiation: approx. 1000 W at +20°C
approx. 200 W at -25°C
Nitrogen inerting with residual oxygen monitoring
With nitrogen inerting, the test chamber is first purged with a large flow rate of N2 until the
o2 content drops below 3%. Then, it switches to a base flow amount (approx. 10% of the large flow rate).
The nitrogen flow for the base amount is measured and displayed using a flowmeter and monitored via a limit switch.
O2 monitoring continuously extracts gas from the test chamber via a flame arrestor and passes it over an O2 sensor which checks the residual oxygen content of the extracted air.
Afterwards, the gas is returned to the test chamber via a second flame arrestor.
The extraction pump is monitored by a volumetric flow meter. Additional details can be found in the Dräger user manual in the appendix.
The gas system is currently deactivated!
All technical data refer to an ambient temperature of +25°C, nominal voltage 400V, without test specimen, without irradiation, and without additional equipment.
Dimensions:
Test chamber volume: 950 liters
Test chamber dimensions: Height approx. 700 mm / Width approx. 900 mm / Depth approx. 1500 mm
External dimensions: Height approx. 2400 mm / Width approx. 1270 mm / Depth approx. 3125 mm
Operating data:
Electrical connection: 400V +6/-10 %, 3/N, 50 Hz
Nominal power: 21.6 kW
Rated current: 36.5 A
Protection class: IP 54
Condensate drain: non-pressurized, R 1/2" inner thread
Sound pressure level:
Oglas je automatski preveden. Moguće su greške u prijevodu.
Dobavljač
Napomena: Registrujte se besplatno ili prijavite, za pristup svim informacijama.
Registriran od: 2012
Pošaljite upit
Telefon & Faks
+49 2861 ... oglasi
Ovi oglasi bi vas također mogli zanimati.
Mali oglas
Tata
417 km
Kondenzator recooler-a
Güntner GVH 090.1.D/5-M(D)680,4 /579,2 kw
Güntner GVH 090.1.D/5-M(D)680,4 /579,2 kw
Mali oglas
Hochdorf
824 km
Jedinica za hlađenje temperature (TCU)
SingleCompact WK3-160-9-50 (IM160F)
SingleCompact WK3-160-9-50 (IM160F)
Mali oglas
Berlin
1.005 km
Aksijalni kondenzator
Güntner AGGVH 100.2A/2x3-J(D)
Güntner AGGVH 100.2A/2x3-J(D)
Mali oglas
Forst
908 km
Taster rezervoar 100 m³, 3 bara, izolovan
Mali oglas
Schmölln
872 km
Hladna soba
Dresdner Kühlanlagenbau
Dresdner Kühlanlagenbau
Mali oglas
Delmenhorst
1.210 km
1000L rezervoara pod pritiskom
10539
10539
Mali oglas
Niedźwiedź
659 km
Dvostepena kompresorska jedinica
BitzerS6G-25.2Y (x4)
BitzerS6G-25.2Y (x4)
Mali oglas
Nattheim
781 km
Stojna pila
Holz-Her1255
Holz-Her1255
Mali oglas
Borken
1.194 km
Klimatska test komora sa vlažnošću 1000 litara
CTSC-40/1000 -40°C bis +180°C
CTSC-40/1000 -40°C bis +180°C
Mali oglas
Wiefelstede
1.253 km
Sistem ventilacije
HeimerWK-61
HeimerWK-61
Vaš oglas je uspješno izbrisan
Došlo je do greške